Rivista di formazione e aggiornamento professionale del pediatra e del medico di base, dal 1982 In collaborazione con l'Associazione Culturale Pediatri

Ricerca

PDF MeBrain

Misurazione non invasiva della bilirubina in Neonatologia: limiti e potenzialità

Non-invasive bilirubin measurement in neonatal care: limits and opportunities

Giulia De Bernardo1, Rosa Teresa Scaramuzzo1, Francesco Gabbriellini1, Barbara Barni1, Ilaria Vannozzi1, Cristina Tuoni1, Francesca Dini1, Massimiliano Ciantelli1, Antonio Boldrini1,2
1UO Neonatologia, Azienda Ospedaliero-Universitaria Pisana, Pisa; 2Università di Pisa


Marzo 2018

Abstract
Background - Besides traditional bilirubin measurement methods, in the past few years transcutaneous measurement has become popular in Neonatal Care Units, as it is easy to use and painless. However, transcutaneous bilirubinometers are currently limited to a screening method for hyperbilirubinemia and are not recommended in newborns exposed to phototherapy.
Objective - The objective of this study is to explore the accuracy of a transcutaneous bilirubinometer by examining its performance both in term and preterm infants for jaundice screening, as well as during phototherapy sessions.
Methods - Model JM-105 Dräger is the bilirubinometer tested in the present research. Transcutaneous and serum measurements were performed simultaneously and eventually analysed.
Results - The bilirubinometer was tested on 116 newborns (gestational age ≥ 26 weeks). 158 paired samples were obtained, 35 of which during phototherapy. Comparison of serum and transcutaneous measurements showed a correlation of 0.85 and 0.89, in term and in preterm infants respectively.
Conclusions - The bilirubinometer JM-105 can be considered a reliable screening tool. Serum confirmation should be obtained if transcutaneous bilirubin levels exceed 11 mg/dl. In neonates exposed to phototherapy, the skin surfaces that are usually unexposed to light, such as the back in case of overhead spotlights and the inguinal region covered by the diaper in case of whole body exposure, may save these patients from frequent blood sampling. However, given the limited number of premature infants and phototherapy patients involved in the present study, the results obtained in these settings should be considered preliminary, and may be eventually verified with a higher number of cases.
Contenuto riservato

Per leggere l'articolo è necessario effettuare il login.

Per l'acquisto del singolo articolo scrivi a abbonamenti@medicoebambino.com (la mail si prepara da sola con gli estremi dell'articolo); se accedi, la richiesta parte con un click.

Bibliografia

  1. American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 2004;114(1):297-316.
  2. Lilien LD, Harris VJ, Ramamurthy RS, Pildes RS. Neonatal osteomyelitis of the calcaneus: complication of heel puncture. J Pediatr 1976;88(3):478-80.
  3. Badiee Z, Mohammadizadeh M, Shamee M. Diagnostic usefulness of transcutaneous bilirubinometry in very preterm newborns. Int J Prev Med 2012;3(4):262-5.
  4. Bosschaart N, Kok JH, Newsum AM, et al. Limitations and opportunities of transcutaneous bilirubin measurements. Pediatrics 2012;129(4):689-94.
  5. Rubaltelli FF, Gourley GR, Loskamp N, et al. Transcutaneous bilirubin measurement: a multicenter evaluation of a new device. Pediatrics 2001;107(6):1264-71.
  6. Bertini G. Iperbilirubinemia neonatale: diagnosi clinico-strumentale. XVI International meeting: Neonatologia e Pediatria alla fine del XX secolo. Conquiste e prospettive. Ancona, 1999.
  7. Hegyi T. Transcutaneous bilirubinometry in the newborn infant: state of the art. J Clin Monit 1986;2(1):53-9.
  8. Società Italiana di Neonatologia, Task force per l’iperbilirubinemia neonatale. Raccomandazioni per il trattamento dell’iperbilirubinemia neonatale. Biomedia 2013;13-14.
  9. Zecca E, Barone G, De Luca D, Marra R, Tiberi E, Romagnoli C. Skin bilirubin measurement during phototherapy in preterm and term newborn infants. Early Hum Dev 2009;85(8):537-40.
  10. Povaluk P, Shwetz AE, Kliemann R. Comparative study between plasma and transcutaneous bilirubin measurements in newborns. Rev Paul Pediatr 2011;29(1):6-12.